Product name: 2018 new Lab2000 single station glove box

    ETEX’s new Lab2000 glove box is a standard 1200 single-station glove box, which can be divided into one-piece and split-type. The vacuum glove box of  this system is specially designed for the study of materials science, chemistry, semiconductor and related industries. Suitable for laser welding. Adopting German BASF deaeration material, American UOP high-efficiency water-absorbing material, the water and oxygen content in the glove box can continuously maintain a high-clean and high-purity gas environment with a gas purity below 1ppm for a long time, hand/automatic control system air pressure; hand/automatic control system The purification status of the gas purification system; automatic control of the reduction process of the gas purification system; automatic prompt and alarm function; system control parameter setting; system parameter recording; system actuator working condition monitoring; transparent front panel makes the operation more convenient and easy, and it is widely used in preparing nano Materials and battery electrode materials.



Box

description

Material: stainless steel 1.4301 (SUS type 304), thickness 3 mm

Size: 1200mm(L) x 750mm(W) x 900mm(H)

volume

0.8 m³

Front window

Panel: 8 mm thick safety tempered glass or 10 mm thick polycarbonate (optional)

Size: 1123mm(L) x 839mm(H)

Glove mouth

Hard aluminum alloy material or polyoxymethylene material (optional)

220 mm diameter, O-ring seal

gloves

Material: Butyl rubber

Thickness: 0.4 mm or 0.8mm (optional)

filter

Outlet and inlet filters, filtration <0.3μm

illumination

LED lighting is located on the top of the front window

Leak rate

Leak detection is carried out in accordance with the oxygen method and pressure change method in ISO 10648-2: 1994 / EJ/T 1096-1999 "Sealed Chamber-Part 2: Tightness Classification and Related Inspection Methods". The final acceptance test is carried out in accordance with the standard of first-level tightness <0.05 vol%/h.

GP20 gas purification system

description

Automatic removal of H2O and O2

Single purification column system, automatic regeneration (optional GP200 double purification column)

Closed gas circulation pipeline

Use voltage

AC 230 V / 50-60 Hz, 10 A or AC 115 V / 50-60 Hz, 20 A (optional)

Working gas

Working gas: N2, Ar, He (purity ≥99.999%)

Regeneration gas: H2 accounts for 5%-10%, the rest is working gas (purity ≥99.999%)

Vacuum pump

Specifications: Rotary vane vacuum pump, equipped with oil mist filter, with gas ballast control

Flow rate: 12 m³/h (7 cfm), double substrate, vacuum degree <2 x 10-3 mbar or dry pump (optional)

Cyclic unit

Oil-free high-speed fan

Air volume: 80 m³/h

valve

KF40 High Vacuum Valve

Leak rate

Leak detection is carried out in accordance with the oxygen method and pressure change method in ISO 10648-2: 1994 / EJ/T 1096-1999 "Sealed Chamber-Part 2: Tightness Classification and Related Inspection Methods". The final acceptance test is carried out according to the standard of level 1 tightness <0.05 vol%/h.

Control cabinet size

600mm (L) x 800mm (D) x 870mm (H)

Transition cabin

Large transition cabin

material

Stainless steel type 304 (1.4301), thickness 3 mm

size

φ360mmx600mm(L) or φ390mmx600mm(L) (optional)

Vacuum

<1 x 10-2 mbar

Small transition cabin

material

Stainless steel type 304 (1.4301), thickness 2 mm

size

φ150mm x 330mm(L)

Vacuum

<1 x 10-2 mbar

Cleaning system

description

After setting the corresponding time and pressure, the system can automatically perform gas replacement on the box.

Analyzer

Oxygen analyzer

size

205 mm (L) x 80 mm (W) x 60 mm (H)

Measuring range

0-1000 ppm

Other oxygen analyzers

GE oxy.IQ™ Oxygen Transmitter

Moisture Analyzer

size

205 mm (L) x 80 mm (W) x 60 mm (H)

Measuring range

0-500 ppm

Other water analyzers

GE VeriDri™ Dew-Point Transmitter

Solvent purification system

description

Material: stainless steel 1.4301 (SUS type 304), thickness 3 mm

Internal size: 220mm(Φ) x 450mm(H)

High quality activated carbon

Optional equipment

refrigerator

Independent control system, integrated in the side panel of the cabinet, temperature -35℃, volume 18L or 32L optional

Main body 304 stainless steel sheet, 5 layers of variable compartments

R404 environmentally friendly refrigerant, imported with original packaging

Heating chamber

Large transition configuration heating system, temperature is 200℃, temperature control is ±1℃

other instructions

certified product

ISO9001 certification, CE certification, UL certification

Warranty

One-year warranty, lifetime maintenance

Application note

For details, please refer to the dangerous, warning, and caution clauses in the manual

Packing size (W×D×H)


weight



navigation

>Home                      >About Us

>Product Center       >Cooperation Case

>News                      > Download Center

>Online Message     >Contact Us

contact

Contact: Mr. Luo

Phone: 400-666-9912

Mobile: 136-9175-1230

Address: 4th Floor, Area C, Building 1, Aishang Technology Industrial Park, No. 108, Honghu Road, Songgang Street, Baoan District, Shenzhen, Guangdong Province


scan

1底部二维码1.jpg

Online

image.png


在线留言1.jpg