Tel: 136-9175-1230
Phone: 0755-23208107
http: www.mingruixiang.cn
Website: en.mingruixiang.cn
Add: 4th Floor, Area C, Building 1, Aishang Technology Industrial Park, No. 108, Honghu Road, Songgang Street, Baoan District, Shenzhen, Guangdong Province
Product name: Internal resistance tester MRX-BT lithium battery internal resistance tester
Index item Index parameter | ||
Measurement methods | Four terminal test method | |
Voltage | Voltage measurement range | 0V~20V |
Precision | ± (0.1% of FS + 0.1% of reading) | |
resistance | Resistance measurement range | 200mΩ2000mΩ two kinds can be automatically switched |
Precision | 0mΩ ~200mΩ ± (0.5% of FS + 0.2% of reading) | |
200mΩ~2000mΩ ± (0.5% of FS + 0.2% of reading) | ||
A/D method | 12-bit SAR successive comparison method | |
frequency | Test frequency | 1kHz |
Sampling Rate | 100 times/S | |
Display refresh rate | 5 times/S | |
Test current | MAX AC 20mA | |
Comparators | Open method | Manually open/close |
Quantity | 10 sets of voltage and internal resistance upper and lower limits | |
Abnormal alarm method | When using the comparator, the buzzer can be set to ON/OFF | |
When voltage or AC internal resistance exceeds the upper and lower limits of the comparator, the buzzer will sound | ||
Green indicator light means qualified, red indicator light means unqualified | ||
Storage function | Can store up to 100,000 sets of test data | |
View function | Can view historical test results | |
Calibration function | Zero point calibration | Calibrate the zero point of voltage and internal resistance at the same time |
Voltage calibration | Use external DC 3V~20V (settable) reference calibration voltage | |
AC internal resistance calibration | Use 10~2000mΩ (settable) precision resistance to calibrate internal resistance | |
External control terminal expansion | Extensible automated testing, expandable USB communication (optional accessories, not included in the standard configuration) | |
Preheat time | More than 3min | |
Equipment working environment requirements | ||
Index item Index parameter | ||
range of working temperature | 0℃~40℃ (within the range of 25±10℃, the measurement accuracy is guaranteed: accuracy drift 0.005% of FS / | |
Storage temperature range | -10℃~50℃ | |
Relative humidity range of working environment | ≤70% RH (no condensation) | |
Storage environment relative humidity range | ≤80% RH (no condensation) | |
Use place | Both indoor and outdoor, easy to carry | |
Fixture specification and equipment size | ||
Index item Index parameter | ||
annex | Polymer clamp or special test lead | |
Charger (5V/1A) | ||
Standard resistance (100mΩ) | ||
USB cable | ||
volume | W*D*H: 120*75*25(mm) | |
weight | About 200g |
Contact: Mr. Luo
Phone: 400-666-9912
Mobile: 136-9175-1230
Address: 4th Floor, Area C, Building 1, Aishang Technology Industrial Park, No. 108, Honghu Road, Songgang Street, Baoan District, Shenzhen, Guangdong Province