Product name: Lab2000 single-station integrated single-sided glove box

    The Lab2000 glove box is a high-performance, high-quality closed cycle working system that automatically absorbs water and oxygen molecules and purifies the working environment. It provides an inert atmosphere of 1ppm O2 and H2O that can meet your specific cleaning requirements. This system is an economical circulating purification system designed to meet customer research and development. 
    The Lab2000 system includes a closed box, a transition chamber, a rotary vane vacuum pump and a circulating purification system integrated with a microcontroller operation panel. The inert gas purification system equipped with the standard Lab2000 system is equipped with a set of purification columns (fully automatic and renewable) to purify and maintain the atmosphere in the glove box.


Box body glove box
descriptionMaterial: stainless steel 1.4301 (SUS type 304), thickness 3 mm
Size: 1200mm(L) x 750mm(W) x 900mm(H)
volume0.8 m³
Front windowPanel: 8 mm thick safety tempered glass or 10 mm thick polycarbonate (optional)
Size: 1123mm(L) x 839mm(H)
Glove mouthHard aluminum alloy material or polyoxymethylene material (optional)
220 mm diameter, O-ring seal
glovesMaterial: Butyl rubber
Thickness: 0.4 mm or 0.8mm (optional)
filterOutlet and inlet filters, filtration <0.3μm
illuminationLED lighting is located on the top of the front window
Leak rateLeak detection is carried out in accordance with the oxygen method and pressure change method in ISO 10648-2: 1994 / EJ/T 1096-1999 "Sealed Chamber-Part 2: Tightness Classification and Related Inspection Methods". The final acceptance test is carried out in accordance with the standard of first-level tightness <0.05 vol%/h.
GP20 gas purification system
descriptionAutomatic removal of H2O and O2
Single purification column system, automatic regeneration (optional GP200 double purification column)
Closed gas circulation pipeline
Use voltageAC 230 V / 50-60 Hz, 10 A or AC 115 V / 50-60 Hz, 20 A (optional)
Working gasWorking gas: N2, Ar, He (purity ≥99.999%)
Regeneration gas: H2 accounts for 5%-10%, the rest is working gas (purity ≥99.999%)
Vacuum pumpSpecifications: Rotary vane vacuum pump, equipped with oil mist filter, with gas ballast control
Flow rate: 12 m³/h (7 cfm), double substrate, vacuum degree <2 x 10-3 mbar or dry pump (optional)
Cyclic unitOil-free high-speed fan
Air volume: 80 m³/h
valveKF40 High Vacuum Valve
Leak rateLeak detection is carried out in accordance with the oxygen method and pressure change method in ISO 10648-2: 1994 / EJ/T 1096-1999 "Sealed Chamber-Part 2: Tightness Classification and Related Inspection Methods". The final acceptance test is carried out according to the standard of level 1 tightness <0.05 vol%/h.
Control cabinet size600mm (L) x 800mm (D) x 870mm (H)
Transition cabin
Large transition cabinmaterialStainless steel type 304 (1.4301), thickness 3 mm
sizeφ360mmx600mm(L) or φ390mmx600mm(L) (optional)
Vacuum<1 x 10-2 mbar
Small transition cabinmaterialStainless steel type 304 (1.4301), thickness 2 mm
sizeφ150mm x 330mm(L)
Vacuum<1 x 10-2 mbar
Cleaning system
descriptionAfter setting the corresponding time and pressure, the system can automatically perform gas replacement on the box.
Analyzer
Oxygen analyzersize205 mm (L) x 80 mm (W) x 60 mm (H)
Measuring range0-1000 ppm
Other oxygen analyzersGE oxy.IQ™ Oxygen Transmitter
Moisture Analyzersize205 mm (L) x 80 mm (W) x 60 mm (H)
Measuring range0-500 ppm
Other water analyzersGE VeriDri™ Dew-Point Transmitter
Solvent purification system
descriptionMaterial: stainless steel 1.4301 (SUS type 304), thickness 3 mm
Internal size: 220mm(Φ) x 450mm(H)
High quality activated carbon
Optional equipment
refrigeratorIndependent control system, integrated in the side panel of the cabinet, temperature -35℃, volume 18L or 32L optional
Main body 304 stainless steel sheet, 5 layers of variable compartments
R404 environmentally friendly refrigerant, imported with original packaging
Heating chamberLarge transition configuration heating system, temperature is 200℃, temperature control is ±1℃
other instructions
certified productISO9001 certification, CE certification, UL certification
WarrantyOne-year warranty, lifetime maintenance
Application noteFor details, please refer to the dangerous, warning, and caution clauses in the manual
Packing size (W×D×H)
weight


Previous:no
navigation

>Home                      >About Us

>Product Center       >Cooperation Case

>News                      > Download Center

>Online Message     >Contact Us

contact

Contact: Mr. Luo

Phone: 400-666-9912

Mobile: 136-9175-1230

Address: 4th Floor, Area C, Building 1, Aishang Technology Industrial Park, No. 108, Honghu Road, Songgang Street, Baoan District, Shenzhen, Guangdong Province


scan

1底部二维码1.jpg

Online

image.png


在线留言1.jpg